Alliance DRx™ 6-Mega Pixel Digital X-Ray System

 
X-Ray Generator

Output Rating

 30/40 kW

kVp Range

 40 to 125 kV, 1kV step

Time Range           

 0.001 to 10 seconds

mA Range

 10 to 400/500 mA

HV Cable

 8 Meters

kVp Accuracy

 <±(1%+1kV)

mAs Accuracy

 <±(3%+0.1mAs)

Generator Base

 Provided

Line Normal, Phase

 230VAC, 1ø, 50/60Hz

 
X-Ray Tube Assembly X-Ray Collimator xray stand
Model/Maker  E7239X         E7242X Control  :Manual, 30 Second timer
Focal Spot Size  1.0/2.0mm    0.6/1.5mm Field Shape  :Rectangular
Max Tube Voltage  125kV Maximum  Field Size  :>43x43cm/ 100cm SID
Max Anode HU  140kHU        200kHU Maximum kV : 150kV
Lamp  :>160 Lux
 
Tube Stand & Table
SID  fixed 1,000mm (40 inch)
Table Top Size  1,400 x 700mm (55.1 x 1.27 inch)
Table Movement  4-way
Tabletop stroke (Longitudinal)  250(±125) mm (9.84(±4.92) inch)
Tabletop stoke (Lateral)  200(±100) mm (7.87(±3.94) inch)
Dimension  1,872(H) x 1,400(W) x 895 (D) mm
 (73.7(H) x 55.1(W) x 35.2 inch)
Weight    180kg (397lb)
   
Digital Image Receptor Imaging Workstation
  Alliance DRx™ 6mp  
Active Image Size  14”x17” (35x43cm) Intel(R) Pentium(R) Duel-Core processor
Image Matrix  6Mega Pixels (Active 2048x2478) Memory:   1GB NECC DDR2 667 MHz SDRAM
Image Resolution  3.1lp/mm OS:   Windows XP Professional
Grid     180 lpi Graphics Card:   256MB ARI RadeonX1300PCI
Dynamic Range  16-bit grayscale